Monaghan 225/SIMV Volume Ventilator - MRI Compatible

Description
The MRI version of the 225/SIMV Ventilator is suitable for ventilating patients in MRI environments up to 1.5 Tesla because virtually all-ferrous material is eliminated. Diagnostic imaging quality can now be maintained during passive ventilation. The MRI version of the 225/SIMV is light- weight, convenient and portable, and shows no attraction to the magnet.

Assembly

  1. Place mainline bacteria filter in filter clamp with flow arrow pointing to the rear of the unit. Tighten clamp.
  2. Place corrugated tubing from outlet port on unit to the inlet port on filter.
  3. DO NOT use heater on this ventilator, as it will bother the MRI results.
  4. Connect end of patient circuit (transport vent circuit) to other end of filter.
  5. DO NOT use support arm in MRI as these have a high metal content.
  6. Place a bacteria filter between the ventilator circuit and the patient; change between patients and PRN.
  7. Wipe down ventilator with disinfectant(Cavacide) between patient uses.

Controls

Volume Operation

  1. Set pressure limit control at a value greater than that expected to be required to deliver the volume.
  2. Using crank set bellows to desired volume.
  3. Adjust flow control.
  4. Set the Respiratory Rate using watch to time the cycles.
  5. Select the desired %O2
  6. Set the desired Mode(Assist, Control, or Assist/Control).
  7. Set trigger sensitivity, if in Assist or Assist/Control modes, as explained above.
  8. Set PEEP level once patient circuit is completely on patient. It is indicated on the patient pressure gauge.
  9. SIMV
    1. SIMV only works in Assist/Control mode.
    2. Set SIMV mode control to 4-16 BPM.
    3. In SIMV, unit does not time cycle to end inspiration because inspiration timer is deactivated.
    4. Adjust Respiratory Rate control for desired SIMV rate.

    For more complete operating instructions or other information refer to the operator's manual found in metal file cabinet in the office.